X-RAY Generators

---X-RAY Generators
X-RAY Generators 2018-04-26T07:11:13+00:00

Bosello HT develops and produces in its facility at Cassano Magnago a wide range of X-Ray mono-block compact generators and High Voltage generators for different purposes, ranging from its industrial X-Ray systems to OEM applications in different fields, such as medical, food inspection, security.

Their highly stable X-Ray output values make Bosello High Voltage devices appreciated for level measurements as well as thickness gauging.

Mono-block Integrated

Technical Data Sheet

Model name XRG 70 IT XRG 110 IT
Application fields Food inspection, level measurements, thickness gauging, radioscopy
Supply mains conditions VAC / Hz 48 VAC A� 5 % max a�� 1.9 ARMS 50/60 Hz 230 VAC A� 10 % max a�� 9 ARMS 50/60 Hz
Output Values
Voltage kV 20 A� 70 (constant potential) 30 A� 100 (constant potential)
kV resolution kV 1
kV accuracy A� 1 %
kV stability A� 0.1 %
Current mA 0.05 A� 0.5 0.5 A� 10
mA resolution mA 0.01 0.1
mA accuracy A� 1 %
mA stability A� 0.1 %
Power W 35 max continuous 1000 max continuous
Radiation coverage 28A� cone beam
Operating values
Generator control by analog signals or digitally by RS 232 digitally by RS 232
Operating frequency kHz 40 (H.V. transformer);

100 (X-Ray tube filament)

25 (H.V. transformer);

25 (X-Ray tube filament)

Operating temperature A�C 10 A� 35
Cooling by air by water, 4 lt/min
X-Ray tube specifications
Focal spot size 0,8 (IEC 336) 1,5 (IEC 336)
Maximum voltage kV 70 120
Maximum power W 110 (continuous) 1600 (continuous)
Total inherent filtration mm 1 (Al equivalent) 1,3 (Al equivalent)
X-Ray leakage A�Sv/h < 1 at 1 m < 2 at 1 m
Dimensions (L x W x H) mm 240 x 153 x 177 580 x 200 x 192
Weight kg 10 40